Previous Page  12 / 66 Next Page
Information
Show Menu
Previous Page 12 / 66 Next Page
Page Background

edfas.org

ELECTRONIC DEVICE FAILURE ANALYSIS | VOLUME 18 NO. 4

12

all six scan chains in the same pack. These six latches are

clocked by the same clock buffer and also are loaded by

the same data buffer during preload mode (Fig. 10). The

chip was subject to LVI analysis. The LVI image, shown in

Fig. 11 with the clock signal in yellow and the data signal

in blue, revealed that the data signal stopped at the latch

callout. Figure 12 is a higher-magnification LVI image at

the latch callout. It further showed that the clock signal is

very weak in the latch callout. However, the clock signal

in the clock buffer is still very strong. Furthermore, the

latches close to the clock buffer have some data signal,

while the latches far fromthe clock buffer do not have any

data signal passing through. This implied that the signal

became degraded between the clock buffer and the latch

callout. Laser voltage probingwas employed to check the

clock signal in the latch callout (Fig. 13). Comparing the

waveform from point 1 to the waveforms from points 2,

3, and 4, it was clearly seen that the waveforms at points

2, 3, and 4 were degraded. Based on layout tracing, the

suspected defect was believed to be a highly resistant via

in the interconnect from the clock buffer to the input of

the latch callout. A subsequent focused ion beam (FIB)

cross section on the suspected highly resistant via found

a hollow V2 (Fig. 14).

SCAN CHAIN SINGLE-LATCH SOFT FAILURE

The third case study is a scan chain single-latch soft

failure. The latch passed at high voltage but failed at low

voltage. Although routine PFA on a scan chain single-latch

hard failure can find a defect in most cases, a soft failure

LVI AND LVP APPLICATIONS IN IN-LINE SCAN CHAIN FAILURE ANALYSIS

(continued from page 10)

Fig. 11

LVI image of the six scan chainswith clock-type failure

in a pack. The red square indicates the latch callout

position.

Fig. 12

Magnified LVI image of Fig. 11 shows a strong clock

signal at the clock buffer and some data signals for

the latches close to the clock buffer, and no data

signals for the latches far from the clock buffer.

Fig. 13

LVP waveforms of the input clock circuitry inside the latch callout show a weak clock signal compared to the one from

the reference latch.