edfas.org ELECTRONIC DEVICE FAILURE ANALYSIS | VOLUME 20 NO. 3 20 Ultrasound frequencies of 40 and 80 MHz were explored. The bias voltage was 500 mV and the modulation fre- quency was 1 kHz. It was discovered that the signal with a 2 µmwiring width could be successfully detected when the carrier frequency was 40 MHz; however, no signal was detectedwith an 80MHz carrier frequency. Figure 6 shows images of samples with dierent resin thicknesses using 40 MHz in carrier frequency. Figure 7 shows the signal- to-noise (S/N) ratio of this measurement. A reduction in S/N ratio occurs with increasing resin thickness; however, it is remarkable that the signal could be detected even with 1000 µm thick resin mold. Figure 8 compares the SOBIRCH and LIT results under the same mold thickness of 320 and 620 µm. The bias voltage was 200 mV and the modulation frequency was 5 Hz in LIT images. Under the influence of the mold’s thermal diusion, the heat point Fig. 6 SOBIRCH imageswith dierentmold resin thickness. Fig. 7 S/N for resin thickness. Fig. 8 SOBIRCH and LIT image of sample with mold resin. Fig. 9 SOBIRCH images with dierent bias voltages. wiring spread to the extent that at 620 µm, localization becomes impossible. Researchers subsequently confirmed the detection sensitivity on dierent bias voltages using the 305 µm thick resin sample. The bias voltage was varied from 10 to 250mV. Figure 9 shows the observed SOBIRCH images. (a) (b) (c) (d) (a) (b) (c) (d) (a) (b) (c) (d)